Tutorial: Valve Calibration

Bpod valve calibration maps opening time (seconds) to dispensed volume (µL) using an exponential fit. Calibration data is stored per-setup in the config-dir YAML.

Prerequisites

Install the calibration optional dependencies (includes both scale drivers):

pip install "murineshiftwork[calibration]"

This pulls in serial-scale-hx711 (Arduino+HX711 custom scale) and serial-scale-bench (RS-232/USB commercial bench scales).

Find the serial path for the scale:

ls -la /dev/serial/by-path/   # identify by physical USB port position
# or
ls /dev/ttyACM*

See setup/SERIAL.md for per-setup port assignments.

Running calibration

msw run -s _test_subject -t _calibration_liquid_dynamic --setup <setup_name> \
    --port-scale /dev/ttyACM1

Follow on-screen prompts: the task opens each valve for a range of durations, the scale reports the dispensed weight, and the fit is written back to the setup YAML.

Static (manual weight entry) variant

msw run -s _test_subject -t _calibration_liquid_static --setup <setup_name>

Enter weights manually when prompted (useful without a connected scale).

Viewing calibration curves

msw calibration plot --setup <setup_name> --out ./calibration_plots/

Plots are saved as PDF (one file per setup, timestamped); R² and fit parameters are printed to stdout.

Setup YAML location

Calibration data is written to <config_dir>/setups/<setup_name>.yaml under calibrations.bpod_valve.<valve_id>:

calibrations:
  bpod_valve:
    1:
      updated: "2026-05-21T10:00:00"
      points: [[20.0, 0.8], [40.0, 1.6], [60.0, 2.4], [80.0, 3.3]]

points is a list of [open_time_ms, volume_ul] pairs collected during the run. The exponential fit is computed at runtime from these points: not stored in the YAML.

Troubleshooting

Symptom Likely cause Fix
R² < 0.95 Air bubbles / inconsistent dispensing Flush, repeat
Scale not found Wrong serial port Check SERIAL.md, try ls /dev/ttyACM*
Curve not monotone Outlier point Remove that point from YAML, re-run
ImportError: serial_scale_hx711 Calibration deps not installed pip install "murineshiftwork[calibration]"